SkyLAB 752

The adventure was never so attractive

5+2 plates

FULLY AUTOMATED

ELISA and IFA

ANALYSER

CE and FDA

APPROVED

skylab

Resources

Pipetting

Washing station

Incubation

Reading

Software

Advantages

Technical Specifications

General specifications
Photometric range
0 – 3.3 OD
Spectral range
400 nm to 700 nm
Filter slots
8
Precision
1 SD (at 0.000 to 0.500 OD)
≤ 1% CV (at 0.501 to 2.000 OD)
≤ 1.5% CV (at 2.001 to 2.500 OD)
≤ 2.0% CV (at > 2.501 OD)
Accuracy
± 0.01 OD or 2.5% (at 0.000 to 3.300 OD)
whichever is greater
Read time
5 seconds, single wavelength
8 seconds, dual wavelength
Kinetic reading
Yes, with embedded result calculation capability
Pipetting specifications
Number of pipettes
2 independent channels using disposable tips for samples and reagents
Sample tip type
Standard disposable tips
Sample tip size
200 –1000 μL
Sample pipetting volume
1 –200 μL (with large tips up to 1000 μL)
Estimated transfer time
< 9 seconds per sample
Dispensing time
< 14 minutes
(for 96 samples with volume of 100 μL)
Precision
Reagents: ≤ 3% CV (10 replicates) for any operating volume above 25 μL
Samples: ≤ 3% CV (10 replicates) for any operating volume above 10 μL
Accuracy
Reagents: ±3% CV (10 replicates) for any operating volume above 25 μL
Samples: ±3% CV (10 replicates) for any operating volume above 10 μL
Adding reagent to full 96 plate
From 3 to 7 minutes depending on reagent volume (50 to 200 μL)
Pipetting scheme for samples and controls
Horizontal or vertical
Dispensing and aspiration speed parameters
Up to 3 different speed profiles for pipetting of samples, controls, diluents and reagents
Faster aspiration / dispensing for non-foamy reagents
Yes
Incubator specifications
Number of incubators
Up to 7, with following configuration:
• 5 independent under plates on the working area
• 1 under plate below the working area
• 1 shaker-incubator unit below the working area
1 to 3 optional extra shaker-incubator units below the working area
Temperature range
RT (+ 7°C) to + 45°C
Accuracy
±1°C
Temperature uniformity
±1°C
Shaking
Yes (optional up to 4 plates) 200 –1000 rpm
Washer specifications
Manifold configuration
1 (8-way wash head)
Programmable volume
50 –2000 μL
Residual wash volume
< 3 μL per well in a flat bottom plate
Wash containers
4 tanks at 2.0 L
Buffer level alarm
Yes (with continuous level sensing)
Waste container
8 L with continuous level sensing
Dispense precision
≤ 5% CV (with 300 μL H2O in 96 well)
Liquid level sensing
Gravimetric for all tanks and waste container
Reader specifications
Photometric range
0 –3.3 OD
Spectral range
400 nm to 700 nm
Filter dots
8
Precision
1 SD (at 0.000 to 0.500 OD)
≤ 1% CV (at 0.501 to 2.000 OD)
≤ 1.5% CV (at 2.001 to 2.500 OD)
≤ 2.0% CV (at > 2.501 OD)
Accuracy
±0.01 OD or 2.5% (at 0.000 to 3.300 OD) whichever is greater
Read time
5 seconds, single wavelength
8 seconds, dual wavelength
Kinetic reading
Yes, with embedded result calculation capability
Dimensions
Width
1300 mm /51.18 in
Depth
940 mm /37.00 in
Height
980 mm /38.58 in
Weight (net)
220 Kg / 485.01 lb
Other specifications
Number of reagents tips
96
Reagent pipetting volume
Up 1000 μL
Number of sample tips loaded
Up to 9 tip racks (864 in total)
No tip number limits in continuous loading mode
Dilution range
1 part in 200 one stage dilution
1 part in 40000 two stages dilution
Direct in-plate predilution
Yes
Barcode reader for samples and resources identification
Yes, with capability of reading EAN/UPC, Code 39, interleave 2 of 5, Code 93 and Code 128 barcode formats
Capability of processing Combi/Profile assays
Yes
Calibration curve data storage
Yes, per any reagent kit lot, with possibility of using stored curves or updating curves using 1 or more points of curve recalibration
Secure access to system functionality
Yes, password protection on system parameters and method programming